    CEMS — Continuous Emission Monitoring Systems

    Category: Emission Monitoring | Gas Analytics | Process Analytical Technology (PAT) | Environmental Measurement
    Author: Dusko Kadijevic, Publisher Analytic Journal
    Last updated: May 2026
    Reading time: approx. 11 minutes
    Target audience: Process engineers, environmental compliance officers, plant operators, instrumentation engineers at permitted industrial facilities


    Definition

    CEMS (Continuous Emission Monitoring System) refers to a system for the continuous, automated measurement of pollutant emissions from industrial facilities — with the objectives of demonstrating compliance with regulatory emission limits in real time, maintaining tamper-proof records, and where applicable feeding data back into process control. CEMS is not voluntary — it is a regulatory obligation whose scope and technical requirements are defined by the operating permit of the individual facility.


    The Starting Point: The Operating Permit

    Before a CEMS system is designed, procured or commissioned, one document determines all subsequent decisions: the operating permit of the facility.

    The regulatory authority specifies in this permit:

    • Which pollutants must be monitored continuously
    • Which limit values must be maintained
    • Which measurement systems are approved for use
    • Which quality assurance procedures must be demonstrated
    • How data must be recorded, stored and transmitted to the authority

    What is measured is therefore not a free decision of the plant operator — it results from a regulatory determination driven by facility type, fuel type, materials processed and local permit conditions.

    Practical note: What is burned or processed in a facility determines directly which pollutants are relevant. Coal-fired power plants must monitor mercury because coal naturally contains it. Waste incineration facilities with mixed waste streams — including older materials, fluorescent lamps, amalgam from dental practices or quecksilver-containing instruments — may also receive mercury-containing inputs. Mercury thermometers from hospitals and medical practices are legally classified as hazardous waste and must be separately collected and disposed of by specialist contractors — they must not enter the general waste stream. In practice, however, it cannot be excluded that mercury-containing materials enter a facility through improper handling of older waste streams. Regulators respond to such operational experience with permanent monitoring requirements.


    Type-Approved Measurement Systems

    In Germany, only type-approved measurement systems (eignungsgeprüfte Messeinrichtungen) may be used for regulatory continuous emission monitoring. Each device must have passed a formal type-approval procedure before it can be used.

    Type approval is conducted according to EN 15267 and national minimum requirements. Results are published by the Federal Environment Agency (UBA) in the Federal Gazette. All systems certified since 2009 are listed at www.qal1.de.

    Equivalent frameworks in other regions:

    RegionApproval FrameworkAuthority
    GermanyEN 15267 / UBA type approvalUmweltbundesamt (UBA)
    EU (other member states)EN 15267 — same norm, national implementation variesNational environmental authorities
    UKMCERTS (Monitoring Certification Scheme)Environment Agency
    USAPerformance Specifications (PS series), 40 CFR Part 60 Appendix BUS EPA

    Practical note — new technologies in the approval process: In one waste incineration facility, halogenated compounds (bromine, chlorine) had been determined by wet-chemical methods for many years. When a laser-based measurement system became available that could perform the same task with greater precision, lower maintenance demand and continuous output, it was still undergoing the type-approval process. The regulatory authority accepted installation of the new system on an agreed basis — provided the wet-chemical reference method continued in parallel until type approval was completed. The new system was technically superior and eventually received full approval. The lesson: regulatory authorities can be pragmatic when the technical superiority is clearly demonstrated and the transition arrangement is transparent.


    Typical Measurement Components in CEMS Installations

    ComponentTypical Measurement PrinciplesTypical Facilities
    SO₂NDIR, UV photometry, FTIRPower plants, cement works, refineries
    NOₓ (NO, NO₂)Chemiluminescence, NDIR, FTIRCombustion plants, engines, gas turbines
    CONDIR, TDLSCombustion plants, coke ovens
    CO₂NDIRPower plants, greenhouse gas reporting
    O₂Paramagnetic, ZrO₂, TDLSCombustion control, all fired installations
    HClFTIR, TDLS, wet-chemicalWaste incineration, chlorine chemistry
    HFFTIR, TDLSGlass industry, aluminium smelters
    Particulate matterScattered light, transmitted light, triboelectricAll fired installations, cement works
    Mercury (Hg)Gold trap technique, CVAAS, CVAF, sorbent methodsCoal and waste power plants, cement
    VOC / TOCFIDCoating plants, solvent processing
    NH₃TDLS, FTIRSCR denitrification systems
    Flue gas flowDifferential pressure, ultrasonic, PitotAll plants requiring mass flow balance

    Mercury Measurement — A Special Case with Dedicated Technology

    Mercury (Hg) is one of the most demanding measurement targets in emission monitoring. It occurs in flue gas in several chemical forms — elemental Hg⁰, divalent Hg²⁺ (reactive) and particle-bound Hg — and requires different measurement principles depending on the regulatory requirement.

    Gold trap technique: Mercury is adsorbed from the flue gas stream onto a gold surface (amalgamation). After a defined collection period, the gold trap is thermally regenerated and the released Hg is detected — typically by cold vapour atomic absorption spectrometry (CVAAS) or atomic fluorescence (CVAF). This method is highly sensitive and type-approved in Germany and recognised in US and EU frameworks.

    Direct continuous measurement systems: Modern CEMS for mercury can measure continuously — either through in-situ laser spectroscopy or through extractive conditioning with subsequent CVAAS/CVAF detection.

    When is mercury monitoring typically required or potentially necessary?

    • Typically required: Coal and lignite power plants — natural mercury content in coal is well-established
    • Potentially required: Waste incineration facilities with mixed or unknown waste streams — depending on waste profile and operational history of the facility
    • Potentially required: Cement works using substitute fuels
    • Site-specific: Facilities processing secondary raw materials or hazardous waste — according to permit and operational history

    Practical note from plant operations: At one incineration facility, mercury appeared in the flue gas even though the operator had not expected any mercury-containing material streams. The regulatory authority subsequently ordered permanent Hg monitoring — as a direct response to this specific operational experience, not because of a general statutory requirement. This illustrates that even when an operator expects no obvious mercury input, a facility can be required to implement permanent Hg monitoring based on its operational history.


    The Emission Data System — Tamper-Proof Core of the CEMS

    A frequently underestimated but regulatorily critical element of CEMS is the emission data acquisition and handling system (in Germany: Emissionsrechner or EDAS; in the US: DAHS — Data Acquisition and Handling System). It receives raw signals from all analysers, calculates emission mass flows and stores data in a tamper-proof manner.

    Why the data system must be directly connected:

    In Germany, the emission data system must be connected directly to the measurement instruments — connection via the process control system (DCS/PLC) is not permitted. The rationale: the DCS is an operational system that could theoretically be manipulated. The emission data system is a separate, tamper-proof unit with clearly regulated data paths and access controls. The same underlying principle — direct, tamper-proof data acquisition without routing through operational control systems — applies in the US (DAHS requirements under 40 CFR Part 75) and in the UK (EA Guidance M2), though the specific technical implementation requirements differ in detail.

    Requirements for the emission data system:

    • Tamper protection — no subsequent modification of stored data possible
    • Audit-trail data storage (minimum 5 years in Germany; similar requirements in US and UK)
    • Documentation of all maintenance interventions and calibrations with timestamps
    • Alarm functions for limit value exceedances
    • Interface for data transmission to the regulatory authority

    Annual verification by independent third parties:

    In addition to the operator's ongoing quality assurance (QAL3 daily checks), an annual verification by an accredited, independent inspection body is required (AST — Annual Surveillance Test under EN 14181). In Germany, accredited bodies such as TÜV and DEKRA frequently carry out this function. In the UK, MCERTS-accredited bodies perform the equivalent role; in the US, independent stack testing firms conduct RATAs (Relative Accuracy Test Audits). These annual checks confirm that the CEMS continues to operate within permitted tolerances under real operating conditions and that QAL3 documentation has been correctly maintained. Results must be submitted to the regulatory authority.


    Quality Assurance — QAL1, QAL2, QAL3 and AST

    The quality assurance framework for CEMS is defined in EN 14181 across the EU and UK. The US equivalent uses Performance Specifications and Appendix F procedures under 40 CFR Part 60.

    LevelDesignationContentWho carries it outFrequency
    QAL1Type approvalDemonstration of fundamental suitability of the measurement systemAccredited testing bodyOnce before commissioning
    QAL2On-site calibrationComparison of CEMS against reference measurement methods under real process conditionsAccredited measurement bodyAt commissioning, then every 3 years
    QAL3Continuous quality controlDaily monitoring of measurement range stability using reference gasesPlant operatorDaily, automated
    ASTAnnual Surveillance TestAnnual review of QAL3 data by independent external bodyAccredited inspection bodyAnnually

    US equivalent: RATA (Relative Accuracy Test Audit) under 40 CFR Part 60, Appendix F — functionally equivalent to QAL2, conducted annually.

    CUSUM charts in QAL3:

    For statistical monitoring of daily calibrations, CUSUM charts (Cumulative Sum Control Charts) are widely used. They detect systematic drift at an early stage — before intervention thresholds in conventional control charts would be triggered. CUSUM charts document not only whether an intervention was necessary but also whether it was statistically justified — important for demonstrating maintenance decisions to the regulatory authority.


    CEMS as a Process Control Tool — Not Just Compliance Monitoring

    Emission monitoring is frequently viewed solely as a compliance instrument. This is too narrow a perspective. CEMS data provide valuable information about the operational condition of the plant itself.

    CO, CO₂, NOₓ and O₂ indicate whether combustion is clean and complete. Changes in these values can signal incorrect air-to-fuel ratio, worn burners, fouling or developing process faults — before limit values are breached or damage occurs. The process can be adjusted faster and energy is not wasted through inefficient combustion.

    Important distinction: In this application, emission measurement is an indirect diagnostic lever — the process is not directly controlled via emission values, but the data serve as diagnostic information for assessing process condition. This is a significant difference from SIL safety applications: in functional safety (SIL), CEMS measurement systems are generally not suitable for direct safety control loops — dedicated, type-approved safety measurement circuits are required for that purpose.

    Practical example with high ROI: At one industrial facility, CO₂ was calculated on the basis of individual annual measurements using conservative flat-rate assumptions that systematically overestimated actual output. After introducing continuous CO₂ monitoring, the actual emission quantity proved substantially lower. This enabled a cause-based calculation instead of a conservative flat-rate estimate — with direct economic consequences for cost allocation. The return on investment of the measurement installation was under 12 months.


    Strategic Consideration: In-House or Outsourced?

    CEMS systems are heavily regulated and subject to frequent normative updates. Operators must track the regulatory landscape continuously. For large facilities or multi-site industrial groups, it can make sense to have the planning, construction and commissioning of CEMS systems carried out by specialist service providers — retaining operational responsibility internally. Specialists in emission measurement technology know the current status of type approvals, accepted systems and the requirements of individual regulatory authorities. Internal competence should then focus on tendering, quality oversight and regulatory communication — not on technical detail planning.


    Regulatory Comparison: Germany, EU, USA, UK

    Germany — BImSchG Framework

    The legal foundation in Germany is the Federal Immission Control Act (BImSchG) and its associated ordinances:

    • 13. BImSchV: Large combustion plants (power stations, combined heat and power)
    • 17. BImSchV: Waste incineration facilities
    • TA Luft (2021): Technical Instructions on Air Quality Control — emission requirements for all permitted facilities

    Type-approved measurement systems are published by the Umweltbundesamt in the Federal Gazette. Certificates are accessible at www.qal1.de.

    EU — Industrial Emissions Directive (IED)

    The Industrial Emissions Directive (IED, 2010/75/EU) sets the European framework, requiring the use of Best Available Techniques (BAT) as defined in BAT Reference Documents (BREFs). Emission limit values follow BAT Conclusions.

    EU-wide for CEMS: EN 14181 for quality assurance of automated measurement systems, EN 15267 for device certification.

    Important note on variation within the EU: While the IED and EN 14181/15267 apply in all member states, the stringency of national implementation and enforcement varies considerably. Germany typically applies requirements stricter than the EU minimum. France, Italy, Spain and Poland implement the same framework but national technical detail requirements and enforcement intensity differ — particularly for older installations which may operate under transition arrangements. Operators at multi-country sites should not assume that German-level requirements apply across all EU locations.

    USA — EPA and Clean Air Act

    In the United States, emission monitoring is governed by the Environmental Protection Agency (EPA) under the Clean Air Act:

    • 40 CFR Part 60 (NSPS): New Source Performance Standards — requirements for new sources
    • 40 CFR Part 63 (NESHAP): National Emission Standards for Hazardous Air Pollutants
    • 40 CFR Part 75: Acid Rain Program — SO₂/NOₓ from electric utility steam generating units
    • 40 CFR Part 98: Greenhouse gas reporting

    Performance Specifications (PS series) define minimum CEMS performance requirements. RATA (Relative Accuracy Test Audit) is the US equivalent of QAL2 — annual comparison against certified reference measurement methods.

    Flares — a US-specific regulatory topic:

    In the United States, flaring of surplus process gases is widespread in petrochemicals, refineries and upstream oil and gas operations. In Germany and the EU, continuous flaring is strongly restricted — surplus gases must be recovered or treated.

    For flares in the US, 40 CFR § 63.670 specifies requirements including:

    • Monitoring of combustion efficiency (Destruction and Removal Efficiency, DRE)
    • Measurement of flue gas velocity and visible emissions
    • Root cause analysis documentation for exceedances
    • For refineries: continuous monitoring of flare gas volume flow and heating value

    CEMS systems at US flares must continuously capture gas flow, composition and heating value to demonstrate combustion efficiency — requirements substantially tightened since 2019.

    UK — Environment Agency and MCERTS

    Following Brexit, the UK applies independent requirements:

    • Environmental Permitting Regulations (EPR) as the permitting framework
    • EA Guidance M2: Monitoring requirements for stack emissions
    • MCERTS (Monitoring Certification Scheme): UK-specific certification programme for CEMS — functionally equivalent to QAL1 in the EU, administered by the Environment Agency
    • EN 14181 and EN 15267 remain the technical reference; MCERTS certification is additionally required for UK regulatory compliance

    Extractive vs. In-Situ CEMS

    CharacteristicExtractive CEMSIn-situ CEMS
    PrincipleSample extracted, conditioned, analysedMeasurement directly in flue gas duct
    Sample conditioningExtensive — coolers, filters, pumps, linesNot required
    Response timeSlower (transport path)Fast (seconds)
    MaintenanceHigh (sampling system)Medium (optics, window cleaning)
    Condensable componentsManageable (heated lines)Conditional
    High particulate loadingFiltration possibleCritical — optical systems sensitive
    Typical principlesNDIR, FTIR, chemiluminescence, gold trapTDLS, UV-DOAS, transmitted light

    Standards and Regulatory References

    Standard / RegulationIssuerContent
    EN 14181CEN / national standards bodiesQuality assurance for automated measuring systems — QAL1, QAL2, QAL3, AST
    EN 15267CENCertification of automated measuring systems for continuous emission monitoring
    EN 17255CENRequirements for data acquisition and handling systems for CEMS
    IED (2010/75/EU)EUIndustrial Emissions Directive — framework for emission limit values and BAT
    40 CFR Part 60US EPANew Source Performance Standards — CEMS requirements for new sources
    40 CFR Part 63US EPANational Emission Standards for Hazardous Air Pollutants
    40 CFR § 63.670US EPASpecific requirements for flare control devices
    40 CFR Part 75US EPAContinuous emission monitoring in the Acid Rain Program
    EA Guidance M2Environment Agency (UK)Monitoring requirements for emissions from stacks
    MCERTSEnvironment Agency (UK)Certification scheme for emission monitoring systems in the UK

    F&A — Frequently Asked Questions

    What is CEMS and why is it mandatory?
    CEMS stands for Continuous Emission Monitoring System — a system for the continuous measurement of pollutant emissions from industrial facilities. It is not a voluntary measure but a regulatory obligation defined in the operating permit of the facility. Which substances must be monitored depends on facility type, fuel and processed materials.

    Who determines what must be measured?
    The regulatory authority — the competent environmental authority in the relevant jurisdiction. In Germany, this is the Länder authority under BImSchG; in the US, EPA and delegated state agencies under the Clean Air Act; in the UK, the Environment Agency under EPR. The operator has no discretion over the basic selection of measurement components.

    What are type-approved measurement systems and where can I find the approved list?
    Devices that have passed a formal type-approval procedure under EN 15267 and been published by the relevant authority. In Germany, the list is maintained by the Umweltbundesamt and accessible at www.qal1.de. In the UK, MCERTS-certified systems are listed by the Environment Agency. In the US, EPA Performance Specifications define minimum requirements rather than individual device approvals.

    Why must the emission data system be directly connected to the analysers?
    The process control system (DCS/PLC) is an operational system that could theoretically be manipulated. A direct connection to the emission data system — without routing through the DCS — ensures tamper-proof data integrity. This principle applies in Germany (EDAS), the US (DAHS under 40 CFR Part 75) and the UK (EA Guidance M2), though specific technical requirements vary.

    What is the difference between QAL2 and RATA?
    Both involve comparison of the CEMS against certified reference measurement methods under real operating conditions. QAL2 is the European designation under EN 14181; RATA (Relative Accuracy Test Audit) is the US designation under 40 CFR Part 60. They are functionally equivalent — the main difference is the regulatory framework and frequency requirements.

    When is mercury monitoring typically required?
    Mercury monitoring is typically required at coal and lignite power plants where mercury content in coal is well-established. It may be required at waste incineration facilities with mixed waste streams, cement works using substitute fuels, and facilities with operational history showing mercury in flue gas. Regulators may impose permanent mercury monitoring based on site-specific experience even when no obvious mercury sources are expected.

    What are CUSUM charts and why are they used in CEMS?
    CUSUM charts (Cumulative Sum Control Charts) are statistical tools for detecting systematic drift in measurement systems. In CEMS operation, they are used for daily QAL3 monitoring. They identify trends early and document statistically whether maintenance interventions were necessary and justified — important for demonstrating maintenance decisions to regulators.

    Why are flares regulated differently in the US compared to Germany and the EU?
    In Germany and the EU, continuous flaring of process gases is strongly restricted — surplus gases must be recovered or treated. In the US, flaring is widespread in petrochemicals, refineries and upstream oil and gas, and is regulated under 40 CFR § 63.670. US flares are subject to comprehensive CEMS requirements for combustion efficiency monitoring, requirements that have been substantially tightened since 2019.

    Does EN 14181 apply equally across all EU member states?
    The norm is the same across all EU member states. However, the stringency of national implementation and enforcement varies considerably. Germany typically applies requirements stricter than the EU minimum. Other member states implement the same framework but with different levels of technical detail and enforcement intensity — particularly for older installations under transition arrangements.

    Can CEMS data be used for process control, not just compliance?
    Yes — and this is an underutilised application. CO, CO₂, NOₓ and O₂ values from CEMS provide direct diagnostic information about combustion quality and process condition. Deviations can signal developing faults before limit values are breached. However, CEMS measurement systems are not suitable for direct safety control loops under SIL functional safety requirements — dedicated, type-approved safety measurement circuits are required for that purpose.


    What This Article Deliberately Does Not Cover

    Sample conditioning systems for extractive CEMS

    Filtration, cooling, heating, condensate removal and pump concepts for extractive CEMS systems are covered in a dedicated article on sample conditioning.

    Further reading: "Sample Conditioning for Process Analysers — Filters, Coolers, Pumps" (forthcoming)


    Dioxin and furan measurement

    Dioxins and furans at waste incineration facilities are measured periodically by accredited testing bodies rather than continuously. This specialist measurement technology is a subject in its own right.


    Share Your Experience

    CEMS is a topic with many practical dimensions — from permit negotiations through system design to QAL3 documentation and the introduction of new measurement principles. If you have experience from plant operations, approval processes or the implementation of new technologies that you would like to share:

    Feedback and practical reports: info@analyticjournal.de

    Analytic Journal aims to be a platform for exchange among practitioners — not a one-way channel. Practical reports from plant operations, corrections and additions are explicitly welcome.


    Summary

    CEMS is not a question of technology selection — it is a regulatory obligation whose framework is defined by the operating permit. What is measured, which systems may be used and how data must be stored and transmitted is prescribed by the authority.

    The technical challenge lies in operating robust, low-maintenance and long-term stable measurement systems within this framework — under the harsh conditions of real flue gas streams, with varying load profiles and under the constant scrutiny of quality assurance requirements.

    New measurement principles — such as TDLS for HCl, NH₃ or mercury — are increasingly receiving type approvals and replacing older, more maintenance-intensive methods. The path to approval of new technologies requires persistence, but regulators are receptive when technical superiority is clearly demonstrated.


    Analytic Journal — The Portal for PAT
    Regulatory requirements are subject to change — current versions should be verified with the relevant authorities.


    © 2026 Analytic Journal / Dusko Kadijevic. All rights reserved.
    This article is protected by copyright. Short excerpts with source attribution and an active link to analyticjournal.com are permitted. Full or substantial reproduction — including in digital form, by AI systems or automated processes — requires the prior written consent of the publisher. Enquiries: info@analyticjournal.de


    Related topics on Analytic Journal:

    • Gas Analytics in the Process Industry — Measurement Principles, Applications and Selection Guide
    • SIL and PFD in PAT — Functional Safety Explained for Practitioners
    • TDLS in PAT — Principles and Applications (forthcoming)
    • Sample Conditioning for Process Analysers — Filters, Coolers, Pumps (forthcoming)
    • Speciality Methods in Gas Analytics — Trace Measurement and Ultra-High Purity Applications (forthcoming)
    • TOC Measurement in Water — Principles, Standards and Manufacturer Overview

    Editorial Note

    This article has been written to the best of our knowledge based on publicly available technical information, standards, and manufacturer data. Specifications may change — for binding information, consult current manufacturer datasheets and relevant standards bodies. Analytic Journal produces editorial content independently of manufacturers and advertising partners.